Surface tension of deionised water, pendant drop
Values reported by the client; water is about 72 mN/m at room temperature
From May 2026 the team's deionised water read about 60 mN/m against about 72 mN/m expected at room temperature, with any needle, drop volume or stabilisation time. A diagnosis call on 26 August 2026 pointed to the needle diameter that sets the image scale. With Droplet Lab's needle size chart and its own in house measurement, the team reported 71 mN/m on 8 October 2026 and closed the issue. Only water has been checked so far.